Model
|
300-C-2
|
300-C-5
|
Cable Length
|
2m
|
5m
|
Conductivity Range
|
μS/cm
0.000 to 0.199
0.200 to 1.999
2.00 to 19.99
20.0 to 199.9
200 to 1999
mS/cm
2.00 to 19.99
20.0 to 199.9
200 to 2000
|
μS/m
0.0 to 19.9
20.0 to 199.9
200 to 1999
mS/m
2.00 to 19.99
20.0 to 199.9
200 to 1999
S/m
2.00 to 19.99
20.0 to 200.0
|
Resolution
|
Auto ranging, up to 4 significant digits
|
Accuracy
|
± 0.5% full scale of each range
> 200 mS/cm (20.0 S/m): ± 1.5% full scale
|
Reference Temperature
|
15 to 30 °C
|
Temperature Coefficient
|
0.00 to 10.00 %/°C
|
Calibration Points
|
Up to 4 (Auto) / Up to 5 (Manual)
|
Units
|
S/cm, S/m
|
Salinity Range
|
0.00 to 80.00 ppt
0.000 to 8.000 %
|
Resolution
|
0.01 ppt, 0.001 %
|
Accuracy
|
± 0.5% of reading or ± 0.01 ppt,
whichever is greater
|
Salinity Curves
|
NaCl, Seawater (UNESCO 1978)
|
Calibration Option
|
Yes
|
Total Dissolved Solids (TDS) Range
|
0.01 mg/L to 200,000 mg/L
|
Resolution
|
0.01 minimum, 4 significant digits
|
Accuracy
|
± 0.5% of reading or ± 0.1 mg/L,
whichever is greater
|
TDS Curves
|
Linear (0.40 to 1.00), EN27888, 442, NaCl
|
Resistivity Range
|
Ω•cm
0.1 to 199.9
200 to 1999
kΩ•cm
2.00 to 19.99
20.0 to 199.9
200 to 1999
MΩ•cm
2.00 to 19.99
20.0 to 200.0
|
Ω•m
0.001 to 1.999
2.00 to 19.99
20.0 to 199.9
200 to 1999
kΩ•m
2.00 to 19.99
20.0 to 199.9
200 to 2000
|
Resolution
|
Auto ranging, up to 4 significant digits
|
Accuracy
|
± 0.5% full scale of each range
> 200 mS/cm (20.0 S/m): ± 1.5% full scale
|
Temperature Range
|
-30.0 to 130.0 °C
-22.0 to 266.0 °F
|
Resolution
|
0.1 °C / °F
|
Accuracy
|
±0.5 °C / ±0.9 °F
|
Calibration Option
|
Yes
|
Body Material
|
ABS and polycarbonate
|
Length & Diameter
|
85 x 30 mm
|
Beoordelingen